Optimization of HPLC Determination of Clotrimazole
Abstract
A HPLC method for the determination of clotrimazole and its degradation products. (2-chlorophenyl)diphenylmethanol and imidazole, in Clotrimazol spray 1 % was developed. The development was focused on the selection of an optimal column, composition and pH of the mobile phase, and internal standard. A Discovery ZR-PBD column (5 ?m, 150 -× 4.6 mm) and spectrophotometric detection at 210 nm were selected. The optimal mobile phase was aqueous acetonitrile 1:1 (pH 9.7). The optimal injection volume and flow-rate were 3 ?l and 1 ml min-1, respectively. On the basis of robustness tests, ibuprofen was used as the internal standard. The analysis time was less than 4.5 min.Downloads
Published
2009-04-15
How to Cite
Žáková, P., Sklenářová, H., Havlíková, L., Matysová, L., & Šatínský, D. (2009). Optimization of HPLC Determination of Clotrimazole. Chemické Listy, 103(3). Retrieved from http://chemicke-listy.cz/ojs3/index.php/chemicke-listy/article/view/1567
Issue
Section
Articles