Determination of Phosphorus by Inductively-Coupled Plasma Mass Spectrometry

Authors

  • Antonín Kaňa Department of Analytical Chemistry, Institute of Chemical Technology, Prague
  • Oto Mestek Department of Analytical Chemistry, Institute of Chemical Technology, Prague

Keywords:

phosphorus, ICP-MS, DRC

Abstract

Determination of P by inductively-coupled plasma MS is complicated by spectral interference of polyatomic ions. In this work the possibility of P determination in biological samples was tested. P was determined as P+ and PO+ ion. Formation of PO+ ions was supported by the reaction with oxygen in a dynamic reaction cell with optimized parameters. The measurements show smaller interference of P+ than that of PO+. The use of a dynamic reaction cell did not lead to improvement of measured parameters. The limit of detection and linearity of calibration curve were determined for this method. Its accuracy was verified by analysis of certified reference skim milk powder and bean samples.

Published

2011-03-15

How to Cite

Kaňa, A., & Mestek, O. (2011). Determination of Phosphorus by Inductively-Coupled Plasma Mass Spectrometry. Chemické Listy, 105(3), 212–216. Retrieved from http://chemicke-listy.cz/ojs3/index.php/chemicke-listy/article/view/1172