Determination of Selenium in Serum Using Atomic Absorption Spectrometry

Autoři

  • M. Ursinyova Institute of Preventive and Clinical Medicine, Bratislava, Slovak Republic
  • V. Hladikova Institute of Preventive and Clinical Medicine, Bratislava, Slovak Republic

Abstrakt

The determination of Se in serum by atomic absorption spectrometry with electrothermal atomization is presented. The method has the detection limit of 12.9 pg Se, which was determined with 100 ml samples. The method was verified using the standard reference material - SeronormTM Trace Element Serum, NYCOMED PHARMA AS, Norway. The median level of Se in the serum of healthy volunteers from Bratislava (Slovakia) was 83.6 mg.l-1). The median concentrations of Se in the serum of male donors were higher (85.7 mg.l-1) than in female donors of the similar age (80.1 mg.l-1).

Publikováno

15.07.1998

Jak citovat

Ursinyova, M., & Hladikova, V. (1998). Determination of Selenium in Serum Using Atomic Absorption Spectrometry. Chemické Listy, 92(6). Získáno z http://chemicke-listy.cz/ojs3/index.php/chemicke-listy/article/view/2698

Číslo

Sekce