Testing the Compliance of Results of Laboratory Determination of C-reactive Protein, Cystatin C, and INR with those Obtained on the Point of Care Testing Analyzer

Page: 190

J. Netriováa,b, Z. Krajčovičováa, V. Meluša, and M. Černickýc

a Faculty of Healthcare, Alexander Dubček University of Trenčín, Trenčín, b St. Michael’s Hospital, Inc., Bratislava, c The part-time doctoral student of St. Elizabeth University of Health and Social Sciences,Bratislava

Keywords: laboratory diagnostics, Passing-Bablok regression, C-reactive protein, cystatin C

 

The use of the Point of Care Test (POCT) analyzers is associated with the requirement of concordance of their results with those obtained by a "classical" way in the medical laboratory. Using the Passing ? Bablok regression method, we verified the compliance of results of testing the parameters of CRP, cystatin C and INR between the AU 400 (Olympus Diagnostica GmbH, D) laboratory analyzer, the Sysmex CA1500 (Siemens, D) laboratory analyzer and the Smart 700 (Eurolyser Diagnostica GmbH, A) POCT analyzer. We found linear dependence of the results of all three tested parameters. Systematic and proportional differences were found only in the case of cystatin C (95% confidence interval for intersection = –0.190; –0.044). On the basis of given data, it can be concluded that an agreement exists between results of methodologically different examinations. For more general acceptance, however, it will be necessary to test a larger number of individuals with a broader range of diagnoses.

 